Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations
2-2 PerSeptive Biosystems
2
2.1 Installing the System
Your Voyager Biospectrometry Workstation is initially installed
by a PerSeptive Biosystems Technical Representative. Do not
use the Voyager system before it is properly installed. Use the
information in this chapter if you move the Voyager system.
2.2 Selecting the Site
This section includes:
Voyager-DE and Voyager-DE PRO Workstations
Voyager-DE STR Workstation
2.2.1 Voyager-DE and
Voyager-DE PRO Workstations
In this section
This section includes:
Required space
Weight
Power/voltage requirements
Selecting input voltage
Required space
The Voyager-DE and Voyager-DE PRO mass spectrometers
measure:
27 inches (69 cm) deep
25 inches (64 cm) wide
65 inches (165 cm) high, includes cabinet and flight tube
Allow an additional 10 inches (26 cm) above the top of the
flight tube for service.
Allow 4 inches (10 cm) at the back of the mass spectrometer
for fans and cables.