Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations
2-34 PerSeptive Biosystems
2
Idle Power Off/Idle Time—When enabled, the
number of minutes after which the high voltage
power supplies automatically turn off, if the
instrument is not used. Default is 60 minutes. If
this value is zero, the high voltage remains on until
any of the following occur: you select
Instrument/Turn off High Voltage, Source Pressure
exceeds Maximum Operating Pressure, you click
Load or Eject, you align a sample plate, or you
exit the software.
4. Click OK to exit.
Timed ion
selector
configuration
To check the Timed Ion Selector (called Precursor Ion Selector
in PSD mode) configuration:
1. In the Instrument Control Panel, select Hardware
Configuration from the Instrument menu.
2. Click the Timed Ion Selector tab to display the Timed Ion
Selector page (Figure 2-18).
Figure 2-18 Timed Ion Selector Configuration