Voyager DP-263-S Guide de l'utilisateur Page 486

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 612
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 485
Appendix A Specifications
A-8 PerSeptive Biosystems
A
Table A-8 Voyager-DE STR Mass Spectrometer Specifications
Condition Specification
Reflector Single-stage with optimized optics for
PSD Analysis
Flight tube (horizontal) Linear mode—2.0 m
Reflector mode—3.0 m
Ion source Two-stage
Ion source voltages Tunable:
Accelerating Voltage—Up to 25,000 V
Grid Voltage—Range determined by
Accelerating Voltage
Laser Nitrogen, 337 nm, 3 ns pulse
Vacuum system Dual differential turbomolecular pumping for
ultra-high vacuum
Multi-vacuum gauge capability for
independent monitoring of source and
analyzer regions
Ion detection Positive and negative
Sample analysis 100-well sample plates
Manual control using control stick or software
Sequence Control software for automated
analysis
Mass range
300,000 Da
Upper limit set by suitably ionized biomolecule.
Vue de la page 485
1 2 ... 481 482 483 484 485 486 487 488 489 490 491 ... 611 612

Commentaires sur ces manuels

Pas de commentaire