Acquiring Spectra from the Sequence Control Panel
Voyager
Biospectrometry
Workstation User’s Guide 6-63
6
How file types
and calibration
specified affect
the data
Figure 6-20 illustrates how these files and selections affect
data.
.CAL Contain calibration constants used
according to Calibration Type:
External calibration
Constants are applied and
saved in .DAT file.
Internal calibration (.CAL
optional)—If .CAL specified,
constants are applied before
the reference masses in the
.SET file are matched.
Constants are updated within
the .DAT file after calibration.
Internal-Update calibration
If .CAL specified exists,
constants are applied before
the reference masses in the
.SET file are matched. (If .CAL
specified does not exist, no
constants are applied.)
Constants are updated within
the .DAT file. Specified .CAL
file is updated or created and
contains new calibration
constants after calibration.
“Creating calibration
(.CAL) files” on
page 6-66
“Calibration File” on
page 6-73
File Type Description For information see