Instrument Settings Parameter Descriptions
Voyager
Biospectrometry
Workstation User’s Guide 5-37
5
Accumulation
Conditions
(continued)
Accumulate all passing (continued)—Example: If
Number to Acquire=5, Number of positions in .SP=7,
the number of positions analyzed is determined by
whether acceptance criteria fail:
If all fail, 7 positions analyzed (total number of positions
in .SP), no .DAT created
If the first position fails, but all remaining pass,
2 positions analyzed, one .DAT created
If the first and second positions fail, but all remaining
pass, 3 positions analyzed, one .DAT created, and
so on
If acceptance criteria never fail, 1 search pattern
position analyzed
For more information, see Section 6.6.6, Process that
Occurs when Accumulating Spectra from Multiple Search
Pattern Positions.
Spectrum Acceptance
Criteria
Click to display Spectrum Acceptance Criteria
parameters. See “Setting spectrum acceptance and laser
adjustment criteria” on page 6-40.
* If Use Automated Sample Positioning is disabled, acquisition is performed on the positions you
manually select.
Table 5-10 Automatic Control Parameters—Spectrum Accumulation (Continued)