Troubleshooting
Voyager
Biospectrometry
Workstation User’s Guide 8-15
8
Dimer in spectrum Laser intensity too
high
Adjust laser by using the
slider controls on the Manual
Laser/Sample Positioning
control page.
Dimers, trimers, and
tetramers in spectrum
Sample
concentration too
high
Prepare sample/matrix with a
final sample concentration
appropriate for sample and
matrix. See Section 3.1.3,
Matrix Information.
Mass calibration on
Spectrum window not
accurate
Wrong Flight Length
entered in Hardware
Configuration
Call PerSeptive Biosystems
Technical Support.
Table 8-1 Spectrum Troubleshooting (Continued)
Symptom Possible Cause Action