Acquiring Spectra from the Sequence Control Panel
Voyager
Biospectrometry
Workstation User’s Guide 6-69
6
NOTE: Run logs are overwritten each time you run a
sequence of the same name. If you want to keep run
logs for future reference, rename the run log each time
you run the sequence.
The run log file contains the list of lines from the run list
that were executed during the sequence, and any
errors that occurred. If a line in the run list generated
more than one data file, log lines are duplicated and file
names are incremented accordingly.
5. Click OK.
6.7.3.2 Creating a Run List
This section includes:
Sample order in the run list
Creating a run list
Modifying and customizing the run list
Importing and exporting the run list
Sample order in
the run list
The type of calibration you perform may require a specific
sample order in the run list:
Internal or external calibration—List samples in any
order.
Close external calibration (internal-update)—List
samples and standards as described in Section 6.7.7.3,
Performing Close External Calibration.
Creating a run list
To create a new run list:
1. Select New Sequence from the File menu.
A new sequence run list grid opens (Figure 6-22).
Vue de la page 380