Chapter 8 Maintenance and Troubleshooting
8-24 PerSeptive Biosystems
8
Sample holder empty
when you click Eject
Sample plate too far way
from grabber in Load
position
Call PerSeptive
Biosystems Technical
Support.
Sample plate jammed in
system
Call PerSeptive
Biosystems Technical
Support.
Table 8-4 Mass Spectrometer Troubleshooting (Continued)
Symptom Possible Cause Action
Table 8-5 Vacuum Gauge Panel Troubleshooting
Symptom Possible Cause Action
E02 error message
displayed on gauge
controller
BA1 or BA2 shut down
due to sudden surge of
high pressure, for
example, when a wet
sample plate is inserted
Press the EMIS button on
Pressure Gauge Control
Panel to turn off. Press
again to turn on.
Call PerSeptive
Biosystems Technical
Support if error occurs
again.
E03 error message
displayed on gauge
controller
Ion gauge connection
broken or disconnected
Call PerSeptive
Biosystems Technical
Support.
E05 error message
displayed on gauge
controller