Voyager DP-263-S Guide de l'utilisateur Page 551

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 612
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 550
Troubleshooting in Continuous Extraction Mode
Voyager
Biospectrometry
Workstation User’s Guide H-23
H
Poor mass accuracy in
Continuous Extraction
mode
(continued)
Incorrect peaks entered in
calibration
Recalibrate. See the
Data
Explorer Software User’s
Guide
, Section 5.3,
Manual Calibration.
(continued on next page)
When calculating peaks,
the Centroid% value
displayed in the Settings
dialog box changed
Use the same Centroid%
value for peak calculation
and calibration (calibration
value is displayed when
you open the Settings
dialog box).
See the
Data Explorer
Software User’s Guide
,
Section 5.5, Centroiding,
for additional information.
Table H-8 Spectrum Troubleshooting (Continued)
Symptom Possible Cause Action
Vue de la page 550
1 2 ... 546 547 548 549 550 551 552 553 554 555 556 ... 611 612

Commentaires sur ces manuels

Pas de commentaire