Chapter 5 Optimizing Instrument Settings
5-76 PerSeptive Biosystems
5
5.4.4.1 Overview
You can set the following parameters to optimize
signal-to-noise ratio:
Accelerating Voltage
Low Mass Gate
Shots per Spectrum
Guide Wire Voltage%
The following sections give guidelines for setting these
parameters.
.
Figure 5-19 Optimizing Signal-to-Noise Ratio
Readjusting the
laser after
optimization
After you have optimized Accelerating Voltage, Guide Wire
Voltage%, and Shots/Spectrum for optimum signal-to-noise,
you can further fine-tune by adjusting the laser intensity. For
information on adjusting the laser intensity, see “Adjusting
laser intensity” on page 5-58.
Signal-to-Noise Ratio
Accelerating Voltage
Guide Wire Voltage%
Shots/Spectrum
Low Mass Gate
To Optimize: Adjust:
Vue de la page 305