Chapter 7 PSD Analysis
7-24 PerSeptive Biosystems
7
7.3.3 Setting PSD Acquisition Parameters
Setting PSD Acquisition parameters includes:
Displaying the PSD Acquisition control page
Setting voltages and external calibration for the precursor
spectrum
Setting precursor mass and PSD calibration for fragment
spectra
Displaying the
PSD Acquisition
control page
To display PSD Acquisition control page:
1. Open the Angiotensin_PSD.BIC file provided. This is
a PSD mode .BIC file.
The PSD Acquisition Settings control page is
automatically displayed if you open a .BIC file that is set
to PSD mode.
To manually display the PSD Acquisition Settings
control page, change the operation mode to PSD by:
Clicking Mode/Digitizer in the Instrument Settings
control page
Clicking the Instrument Mode tab
Selecting PSD
NOTE: Low Mass Gate is automatically disabled if PSD
is specified for Instrument Mode. The Precursor Ion
Selector provides the selectivity needed to screen out
unwanted masses.